Home > Water Treatment > Laboratory Products

HI 221

pH Meters with Calibration Check

  • Description
  • Specifications
  • Included Reagents
  • Accessories
  • Order Info
  • Downloads

A pH electrode that is properly manufactured and kept clean will retain its measuring integrity for quite a long time.

The most common cause for pH measurement inaccuracies is an unclean or improperly cleaned electrode. This is very important to note because, during calibration, the instrument assumes that the electrode is clean and that the standardization curve created during the calibration process will remain a valid reference until the next calibration.

pH meters on the market today will allow an offset of approximately ±60 mV. The deviation from 0 mV is not unusual, in fact it represents the true characteristics of the pH electrode. The deviation from 0 mV becomes a problem if it is the result of calibrating a dirty electrode.

HANNA®' HI 221 & HI 223 compare the characteristics of the pH electrode from one calibration to the next. In case of large variances in the electrode condition that can only be the result of a soiled sensor, these meters alert the user that the electrode needs to be properly cleaned prior to calibration and measuring.

Range pH -2.00 to 16.00 pH
Range mV ±699.9 mV; ±2000 mV
Range Temperature -20.0 to 120.0°C
Resolution pH 0.01 pH
Resolution mV 0.1 (±699.9 mV); 1 (±2000 mV)
Resolution Temperature 0.1°C
Accuracy pH ±0.01 pH
Accuracy mV ±0.2 mV; ±1 mV
Accuracy Temperature ±0.4°C
Calibration Check   status of electrode condition and response time,
status of the buffer solution during calibration
pH Calibration   automatic, up to 5 points with 7 memorized values
(pH 1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45)
Temperature Compensation   manual or automatic, -20.0 to -120.0°C (-4 to 248°F)
pH Electrode   HI 1131P, glass-body, single junction, refillable,
BNC + Pin connector, 1 m (3.3') cable (included)
Temperature Probe   HI 7669/2W with 1 m (3.3') cable (included)
PC Connection   RS232 opto-isolated serial port
Data Logging   100 samples
Input Impedance   1012 Ohm
Power Supply   12 Vdc adapter (included)
Environment   0 to 50°C (32 to 122°F); RH max 95%
Dimensions   240 x 182 x 74 mm (9.4 x 7.2 x 2.9")
Weight   1.1 kg (2.4 lb.)
Electrodes
HI 105X Series Laboratory Combination pH and ORP Electrodes
HI 1083 Series Laboratory Combination pH and ORP Electrodes
HI 13X2 Series Rugged pH and ORP Electrodes for General Use
Solutions
HI 70004P 4.01 pH Buffer Solution
25 x 20 mL sachets
HI 70007P 7.01 pH Buffer Solution
25 x 20 mL sachets
HI 70010P 10.01 pH Buffer Solution
25 x 20 mL sachets
HI 7001L 1.68 pH Buffer Solution
1 x 500 mL bottle
HI 7004L 4.01 pH Buffer Solution
1 x 500 mL bottle
HI 7004/1L 4.01 pH Buffer Solution
1 x 1 L bottle
HI 7004/1G 4.01 pH Buffer Solution
1 x 1 Gallon (3.78 L) bottle
HI 7006L 6.86 pH Buffer Solution
1 x 500 mL bottle
HI 7006/1L 6.86 pH Buffer Solution
1 x 1 L bottle
HI 7006/1G 6.86 pH Buffer Solution
1 x 1 Gallon (3.78 L) bottle
HI 7007L 7.01 pH Buffer Solution
1 x 500 mL bottle
HI 7007/1L 7.01 pH Buffer Solution
1 x 1 L bottle
HI 7007/1G 7.01 pH Buffer Solution
1 x 1 Gallon (3.78 L) bottle
HI 7009L 9.18 pH Buffer Solution
1 x 500 mL bottle
HI 7009/1L 9.18 pH Buffer Solution
1 x 1 L bottle
HI 7009/1G 9.18 pH Buffer Solution
1 x 1 Gallon (3.78 L) bottle
HI 7010L 10.01 pH Buffer Solution
1 x 500 mL bottle
HI 7010/1L 10.01 pH Buffer Solution
1 x 1 L bottle
HI 7010/1G 10.01 pH Buffer Solution
1 x 1 Gallon (3.78 L) bottle
HI 70300L Electrode Storage Solution
1 x 500 mL bottle
HI 70000P Rinsing Solution 25 x 20 mL sachets
HI 7061L General Purpose Cleaning Solution
1 x 500 mL bottle
HI 7073L Proteins
1 x 500 mL bottle
HI 7074L Inorganic Substances
1 x 500 mL bottle
HI 7077L Oil and Fats
1 x 500 mL bottle
HI 7071 Electrolyte Solution, 3.5M KCl + AgCl
4 x 30 mL bottles
HI 7072 Electrolyte Solution, 1M KNO³
4 x 30 mL bottles
HI 7082 Electrolyte Solution, 3.5M KCl
4 x 30 mL bottle
Accessories
HI 920010 Cable for PC Connection, 9 to 9-pin
HI 92000 Windows® Compatible Software for Logging Instruments,
Photometers and Turbidity Meters
HI 221 is supplied with HI 1131P refillable pH electrode (glass-body, BNC & Pin with 1m cable), HI 7669/2W stainless steel temperature probe, HI 76404 electrode holder, pH 4 and pH 7 buffer sachets, electrode refilling solution, 12 Vdc power adapter and instructions.

Manual: (Download)