| Search By Accuracy | ||
| Parameter: | Accuracy: | |
|
|
||
Home > Mini Panel Mounted Controllers > Mini Panel Mounted Controllers
BL 983324 is a panel mounted TDS controller, designed for easy installation, configuration and maintenance.
The meter is provided with a dosing relay, activated when the TDS reading exceeds the set point value.
Measurements are compensated for temperature variations and shown on the display automatically.
A multicolored LED on the front panel continuously indicates if the mini controller is in measurement, dosing or alarm mode.
Wiring and external device connections are extremely simple to perform through the terminals on the rear of the instrument.
Users can choose automatic or manual dosing mode with a switch on the front panel. Manual control is particularly useful during maintenance operations, because it permits operators to enable or disable the dosing relay according to need.
The overtime control system advises operators when the relay is active too long, helping to prevent over dosage.
Order Information:
BL 983324-0 (12 VDC) and BL 983324-1 ( 115/230V) are supplied with mounting brackets, transparent cover and instruction manual.

| Range | 0.0 to 49.9 ppm | |
| Resolution | 0.1 ppm | |
| Accuracy @ 25°C | ±2% F.S. | |
| TDS Conversion Factor | 0.5 | |
| Probe | HI 7634-00 EC/TDS probe with internal temperature sensor and 2 m (6.6’) cable (not included) | |
| Temperature Compensation | Automatic, 5 to 50°C (41 to 122°F) with ß = 2%/°C | |
| Calibration | manual, with CAL trimmer | |
| Dosing Relay | maximum 2A (fuse protected), 250 Vac, 30 VDC contact close when measure > set point | |
| Setpoint | adjustable from 0 to 49.9 ppm | |
| Overtime | adjustable, typically from 5 to approx. 30 minutes | |
| Power Supply | See Chart | |
| Dimensions | 79 x 49 x 95 mm (3.1 x 1.9 x 3.7’’) | |
| Weight | See Chart |
| Electrodes | |
|---|---|
| HI 7634-00 | TDS Probe, 2 m cable |
| Solutions | |
| HI 7061L | General Purpose Cleaning Solution 1 x 500 mL bottle |
| HI 7033L | 84 µS/cm EC Solution, 500 mL Bottle |
| Accessories | |
| HI 710005 | Power Adapter, 115 Vac to 12 VDC, US Plug |
| HI 710006 | Power Adapter, 230 Vac to 12 VDC, European Plug |
| HI 740146 | Mounting Brackets for BL 931700 Housing |
| Probes | |
| HI 7634-00/4 | TDS Probe, 4 m cable |